Pressure measurements in a new TLSO with a dynamic thoracic brace pad for idiopathic scoliosis
Scoliosis volume 4, Article number: O42 (2009)
Introduction
Van den Hout (2002) [1] performed pressure measurements which are practicable and of value for studies on the working mechanism of brace treatment. Therefore, we performed pressure measurements in our new TLSO with a more or less dynamic thoracic brace pad.
Materials and methods
Pressure measurements were performed in 8 patients who were treated with our new TLSO. These were done in 15 positions which can be compared with all day activity.
Results
Eight girls, with a mean age of 12.7 years were asked to participate. The mean duration of brace treatment prior to participating in the present study was 7 months. The mean primary right thoracic curve was 31° and the mean secondary curve measured 23°. The mean corrective force over the lumbar brace pad in standing position was 71 N; over the thoracic brace pad it was 107 N. Van den Hout found in the Boston brace that the mean corrective force through the lumbar brace pad was larger than the mean corrective force over the thoracic brace pad. However, we found higher corrective forces in the thoracic brace pad than in the lumbar brace pad.
Conclusion
Our new TLSO results in higher corrective force measures than in the lumbar brace pad which suggest that the thoracic brace pad is more or less dynamic.
References
Hout van den JA, van Rhijn LW, Munckhof van den RJ, van Ooy A: Interface corrective force measurements in Boston brace treatment. Eur Spine J. 2002, 11 (4): 332-5. 10.1007/s00586-001-0379-1.
Veldhuizen AG: Idiopathic scoliosis. Thesis. 1985, University of Groningen
Andriacchi TP, Schultz AB, Belytschko TB, Dewald R: Milwaukee brace correction of idiopathic scoliosis. A biomechanical analysis and a restrospective study. J Bone Joint Surg Am. 1976, 58 (6): 806-15.
Author information
Authors and Affiliations
Rights and permissions
Open Access This article is published under license to BioMed Central Ltd. This is an Open Access article is distributed under the terms of the Creative Commons Attribution 2.0 International License (https://creativecommons.org/licenses/by/2.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
About this article
Cite this article
Hermus, J., Monteban, P., Guldemond, N. et al. Pressure measurements in a new TLSO with a dynamic thoracic brace pad for idiopathic scoliosis. Scoliosis 4 (Suppl 1), O42 (2009). https://doi.org/10.1186/1748-7161-4-S1-O42
Published:
DOI: https://doi.org/10.1186/1748-7161-4-S1-O42