In-brace correction of the Cobb angle with RSC-CAD CAM compared with 'hand made' from the original author

Introduction

The purpose of the present study is the comparison of the in-brace correction in a group of patients with idiopathic scoliosis treated with ‘CAD CAM’ RSC brace and a control group of patients treated with a ‘hand made’ RSC brace (from the original author, Rigo).

Background

RSC braces built with informatics assistance (CAD CAM) have shown similar in-brace corrections than ‘hand made’ RSC; however this has never been compared in matched samples.

Method

Case control study. Source: Ortholutions database from 2008 to 2009. Inclusion criteria: Patients with idiopathic scoliosis, no previous treatment (brace), younger than 15 and Risser 3 or less. Case group: 41 patients (2 males and 39 females) treated with CAD CAM RSC brace full filed the inclusion criteria. Control group: all the 27 patients (1 male 26 females) treated with ‘hand made’ RSC brace that full filed the inclusion criteria as well. Mean age was not different in the case group 11.3 y (SD 1.9) compared with the control 11.8 y (SD 2.1). Risser sign was higher (p< 0.005) in the case group 0.8 (SD 1) in comparison with the control group 0.2 (SD 0.6). The Cobb angle out-brace of the major curve was not significantly different in both groups (case= 35.2° + 13°; control= 33° + 7.7°). Minor curve was not different (case= 26.4° +/- 11.8°; control=28.8° +/- 7.2°).

Results

The in-brace Cobb angle of the major curve was 18.6°+/-14.1° in the case group and 16.4° +/- 9.9° in the control (NS). Minor curve measured 15.5° +/- 12.9° in the case group and 18.4° +/- 6.9° in the control (NS). In-brace correction of the major curve (case 52.6%; control 53.7%) and the minor curve (case 45.6%; control 36.8%) was not significantly different in both groups. Total in-brace correction (major and minor curve) was also the same in both groups (case 49.3%; control 47.9%).

Discussion

The results of this study show that the Cobb angle can be corrected in the same proportion by using a CAD CAM brace in comparison with a brace built on a positive plaster mould modified by the original author of the RSC brace (Rigo). The CAD CAM ortholutions system uses a highly complete library of moulds provided by the original author of the RSC brace. Brace model is selected from the library for each patient according to his/her curve pattern (Rigo Classification) and morphology by the author. Afterwards, the brace is built by Ortholutions using CAD CAM according to specific patient measurements and fitted by an experienced and specifically trained orthopaedic technician. Both, patient measurement and fitting, are essential parts of the process in order to reach a proper treatment quality. When the full protocol is followed the CAD CAM brace is as effective and safe in reducing the Cobb angle in brace in the major as well as in the minor curve. In this present

study, the CAD CAM RSC brace reduced even more the minor curve in comparison with the hand made, although difference was not statistically significant, most probably due to the sample size. However, in-brace angles of the major and minor curve were more balanced with the hand made brace. We interpret this as a sign of the special purpose of the original author to reach such an in-brace correction balance, when fitting, rather than real structural differences between the two methods.

Conclusion

The in-brace correction of the Cobb angle achieved with a CAD CAM RSC brace is as effective as that observed with a ‘hand made’ brace from the original author.

Author information

Authors and Affiliations

Authors

Rights and permissions

Open Access This article is published under license to BioMed Central Ltd. This is an Open Access article is distributed under the terms of the Creative Commons Attribution 2.0 International License (https://creativecommons.org/licenses/by/2.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.

Reprints and permissions

About this article

Cite this article

Rigo, M.D., Gallo, D. & Dallmayer, R. In-brace correction of the Cobb angle with RSC-CAD CAM compared with 'hand made' from the original author. Scoliosis 5 (Suppl 1), O68 (2010). https://doi.org/10.1186/1748-7161-5-S1-O68

Download citation

  • Published:

  • DOI: https://doi.org/10.1186/1748-7161-5-S1-O68

Keywords