Introduction of a novel optical measurement system for dynamic spinal curvature analysis
Scoliosis volume 8, Article number: O31 (2013)
Background
Rasterstereography is a method for stereophotogrammetric surface measurements of the back based on moiré topography[1]. In contrast to all other optical devices, rasterstereography allows an analysis not only of the back surface but also of the underlying spine. This is possible by the use of a spine model, which was created by Turner-Smith[2, 3]. So far rasterstereography only allows static measurements. However, it would be of great value to be able to also evaluate the spine and its segments under dynamic conditions.
Purpose
The purpose of this study was to evaluate the accuracy of the marker detection of a novel spine and surface topography system that allows measurements under dynamic conditions.
Methods
A new rasterstereographic device (Diers, Germany) was evaluated, by comparison with the gold standard in motion analysis the VICON system. 12 flat infrared markers were adhered to a wooden plate and on the backs of 8 test subjects. By linking the markers, 4 triangles were defined, and the sides of each triangle were acquired under static and dynamic conditions while subjects walked on a treadmill.
Results
On the wooden plate the sides of the 4 triangles were measured by hand with a measuring tape and by the two optical systems. Here, the rasterstereograph showed a higher accuracy in marker detection than the VICON system (0.13 ± 0.84mm vs. 0.73 ±0.71mm). Under dynamic conditions when the subjects walked on a treadmill, the rasterstereographically-measured sides of all four triangles were compared with the sides measured by the VICON system. No significant differences (p>0.05) were found in the marker detection between the systems, as the marker detection differed between 0.07 - 1.1% for all sides of the four triangles.
Conclusions and discussion
The accuracy of the rasterstereograph is superior on a wooden plate and comparable during dynamic measurements to the VICON system, with the advantage that it calculates a three-dimensional surface map and also allows the analysis of the underlying spine. The measurements made under dynamic conditions can be factored into the spinal model to allow an analysis of the scoliotic spine during motion.
References
Willner S: Moire topography for the diagnosis and documentation of scoliosis. Acta Orthop Scand. 1979, 295-302.
Turner-Smith AR: A television/computer three-dimensional surface shape measurement system. J Biomech. 1988, 515-529.
Turner-Smith S: A method for analysis of back shape in scoliosis. J Biomech. 2003, 497-509.
Author information
Authors and Affiliations
Corresponding author
Rights and permissions
This article is published under license to BioMed Central Ltd. This is an Open Access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/2.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
About this article
Cite this article
Betsch, M., Rappr, W. Introduction of a novel optical measurement system for dynamic spinal curvature analysis. Scoliosis 8 (Suppl 2), O31 (2013). https://doi.org/10.1186/1748-7161-8-S2-O31
Published:
DOI: https://doi.org/10.1186/1748-7161-8-S2-O31
Keywords
- Surface Topography
- Dynamic Condition
- Motion Analysis
- Test Subject
- Measuring Tape